(See Table 6.)Ĭontrol and linearity test kits (if available): Accu-Chek Performa control solution-Refer to the control solution package insert for details.Īccu-Chek linearity test kit-Refer to the linearity test kit package insert for details.Īccu-Chek Performa: Healthcare Professional Information: Important information: These test strips are labeled with a green This reaction creates a harmless DC electrical current that the meter interprets for the blood glucose result The sample and environmental conditions are evaluated using AC and DC signals. Q-GDH) from Acinetobacter calcoaceticus, recombinant in E.coli, converts the glucose in the blood sample to gluconolactone. Test principle: The enzyme on the test strip, mutant variant of quinoprotein glucose dehydrogenase (Mut. Performance assessment by the user: A study evaluating glucose values from fingertip capillary blood samples obtained by 209 lay persons showed the following results: For glucose concentrations less than 100 mg/dL (less than 5.55 mmol/L), 97.6 % of the test results were within ☑5 mg/dL (within ☐.83 mmol/L) of the results obtained through laboratory testing.įor glucose concentrations equal to or greater than 100 mg/dL (equal to or greater than 5.55 mmol/L), 97.0 % of the test results were within ☑5 % of the results obtained through laboratory testing. System Accuracy: See Tables 1,2,3,4 and 5.